Design of titanium uncemented femoral stems for hip prosthesis suitable for the Colombian young adult population
Item Links
URI: http://hdl.handle.net/10818/59917Visitar enlace: https://www.scopus.com/inward/ ...
ISSN: 10255842
DOI: 10.1080/10255842.2023.2205978
Compartir
Statistics
View Usage StatisticsBibliographic cataloging
Show full item recordDate
2023Abstract
The increase of revision surgeries in hip replacement procedure in Colombian young adult population can be addressed by a new design of femoral stem that reduces stress shielding. A new femoral stem was designed using topology optimization as a design aid to reduce the mass in the femoral stem and its overall stiffness, combined with the theoretical, computational, and experimental assessment of the new design that complies with a static and fatigue safety factor greater than one. The new femoral stem design can be used as a design tool to reduce the number of revision surgeries caused by stress shielding. © 2023 Informa UK Limited, trading as Taylor & Francis Group.
Ubication
Computer Methods in Biomechanics and Biomedical Engineering
Collections to which it belong
- Facultad de Ingeniería [506]